noise_prewhitening_scale_factor ​

Use this correction only when acquisition and noise-prescan receiver conditions are known. The dwell-time ratio does not by itself identify a scanner-specific effective bandwidth correction. See the noise model.

julia
using HighOrderMRI
scale = noise_prewhitening_scale_factor(4e-6, 8e-6;
    receiver_bandwidth_ratio=1)
@assert scale == 0.5

This is arithmetic with illustrative inputs, not a calibration recommendation.

API ​

julia
noise_prewhitening_scale_factor(acquisition_dwell_time::Real,
    noise_dwell_time::Real;
    receiver_bandwidth_ratio::Real = 1)

Compute the dimensionless correction used when a noise prescan and the MRI acquisition have different sampling intervals or effective receiver bandwidths:

s=TacqTnoiseRBW.

Here receiver_bandwidth_ratio is the scanner- or receiver-specific effective noise-bandwidth correction. Both dwell times must use the same units. Pass the result as prewhitening_scale_factor to fit_coil_compression or compress_coils. A value of one means no correction.

With a covariance Ψ_noise estimated from the noise prescan, the corrected whitening treats Ψ_acq = Ψ_noise / s as the covariance of the acquired MRI data. This follows the scale-factor convention used by ISMRMRD and mrpro; it deliberately does not add an extra sqrt(2) factor because this package defines covariance directly as E[n' * n] for complex noise.

Source: src/Recon/CoilCompression.jl