Reconstruction conventions
Use the same conventions throughout data preparation, operator construction and reconstruction. This page is a practical checklist for comparing implementations; it does not prescribe a particular experiment or report dataset results. Detailed definitions are in Symbols and notation and the encoding model.
Encoding model
The package uses positive forward phase, in cycles before multiplication by
LowRank approximates the residual encoding while retaining Fourier encoding on its active NFFT axes. In 2D, the remaining
Optional concomitant coefficients must be time-integrated and use the documented basis and units. recon_terms selects the original solid-harmonic rows; it does not independently switch concomitant contributions.
Spatial grid
Use coordinates, field maps, masks and coil sensitivities in the same frame and physical units. The default axis
Odd grids include zero; even grids use symmetric half-integer offsets. LowRank assumes its documented Cartesian lattice on NFFT axes. Editing a coordinate vector alone does not change the Fourier plan. Geometry export and resampling have additional implementation limits.
Solid-harmonic term order
Keep kspha rows in the documented order. A coefficient times its spatial polynomial must give phase in cycles. The first row is spatially constant; rows 2–4 are x, y and z. Fieldmap values in Hz require time values in seconds. No implicit gyromagnetic-ratio conversion should be assumed when converting gradient waveforms.
Input shapes and vector ordering
| Operator/input | Layout |
|---|---|
Explicit kspha, times | (nTerm,nSam), (nSam,) |
LowRank kspha, times | (nTerm,nSam,nDyn), (nSam,nDyn) |
| LowRank data | (nSam,nDyn,nCha) |
| LowRank vector output | Samples vary fastest, then readouts, then coils. |
Here nDyn is the code's readout count, indexed by
Reconstruction-term selection
For nine coefficient rows, the three digits of recon_terms select zeroth, first and second order. Sixteen rows use four digits, adding third order. In the shared preparation path, disabled first-order measured coefficients are replaced by the supplied nominal trajectory; disabled zeroth and higher-order rows are zeroed. Check each constructor's actual default and nominal-trajectory requirements.
NFFT convention
The physical Fourier coefficient is scaled by voxel spacing and negated to match the package's positive-phase model through the NFFT's negative-exponent convention. Even-axis centring corrections are part of the implementation. Use the encoding derivation, not a second hand-maintained phase convention.
Weights and regularization
recon_HOOp takes raw sample weights
with normal samplingDensity returns square-root density weights. Do not square these again before passing them to the reconstruction wrapper. Direct optimized normal interfaces have their own documented squared-diagonal contract.
The unregularized normal is positive semidefinite; positive definiteness and a unique solution require additional rank/regularization conditions. Fix the regularizer and inner solver settings before comparing operators. Keep image-domain phase or intensity alignment separate from raw complex reconstruction error.
Reproducibility and interpretation
Record the code revision, data preparation, units, coordinate transformations, mask, coil transform, weights, solver settings and LowRank approximation parameters. Apply a fitted coil transform consistently to data and CSM. A lower residual or a visually different image does not independently prove a more accurate physical model; use the validation guide.