Field preprocessing and synchronization ​

Field coefficients must be aligned with ADC samples and use the encoding operator's phase, coordinate and unit conventions. HighOrderMRI provides GIRF prediction, time interpolation and model-based delay estimation. Field monitoring provides measured coefficients; it does not remove the need to check these conventions. [6]

GIRF prediction ​

A gradient impulse response function models the gradient chain as a linear, time-invariant system. A calibrated frequency response predicts field waveforms from prescribed physical gradients; the approximation and calibration conditions come from the measurement, not the container type. [7]

GIRFModel stores a centred transfer matrix and frequency metadata. apply_girf applies that matrix to three DCS gradient channels. Extra dimensions, such as shots, are evaluated independently. Its real output has the calibrated number of field channels.

The identity-response example is a complete CPU configuration. A real acquisition requires a response matched to the waveform dwell time and calibrated channel units. The model's frequency vector is not used for automatic resampling. The current FFT operation has periodic boundary behavior, and response expansion can omit a valid impulse coefficient; see current limits.

To convert instantaneous frequency-field coefficients into accumulated phase, use trapezoidal integration with grad2traj. This helper does not supply a gyromagnetic ratio or a 2π conversion. For example, after converting a physical first-order gradient to Hz/m, integration in seconds gives cycles/m. A zeroth-order field in Hz integrates to cycles. Conversion factors must match each output channel's physical meaning.

Interpolation to ADC times ​

InterpTrajTime accepts coefficient samples in rows and field terms in columns. For shot i, field sample j is placed at

tj,i=(j−1)dt−(StartTimei+τ).

Evaluating at the ADC time therefore samples the original field as ki(tADC+StartTimei+τ). This fixes the sign convention: a positive offset queries a later point in the original waveform. Do not describe it as moving the input support to a later time.

Use the linear-signal example to check this sign. Transpose its two-dimensional result before passing it to an explicit encoding constructor, which expects terms in rows. Multishot interpolation uses one time row and one offset per shot. Zero extrapolation outside the measured interval can distort the model; inspect the complete shifted ADC interval before interpreting reconstructed results.

Model-based delay estimation ​

FindDelay and FindDelay_multishot follow the model-based synchronization approach of Dubovan and Baron. [8] The package uses density weights in each image reconstruction, then an unweighted scalar delay update with JumpFact acceleration.

At a current delay τ, reconstruct xτ using Aτ, and define

rτ=y−Aτxτ,bτ=Bτxτ.

For nonzero derivative information, the linearization rτ≈Δτbτ gives the implemented update

Δτ=JRe(bτHrτbτHbτ).

Here J is the current JumpFact. Derivative-forward coefficients are formed with the source's five-tap temporal filter and interpolated using the same delay. For the positive phase convention,

∂∂τei2πΦτ=ei2πΦτi2π∂Φτ∂τ.

The implementation differentiates shifted field coefficients; it does not shift the ADC times multiplying the static field map. After updating τ, J is reduced when the update changes sign. The stopping comparison uses the update magnitude, and the scalar returned is τ−Δτ(J−1)/J with the final J. These implementation choices are not an assertion of exact equivalence to every detail of the publication.

Use the CPU configuration on the single-shot API page: arrayType=Array, a solver type such as CGNR, and a regularization object. There are no use_gpu or λ keywords. Use seconds when the field map is in Hz, and provide Δτ_min explicitly with the input floating-point type. The actual default 0.005 means 5 ms in that convention; it is not 0.005 microseconds. The log prints τ/dt and Δτ/dt, which are sample offsets despite their labels.

Current limits of the delay estimators

There is no outer-iteration cap or convergence-status result. iter_max only limits an image reconstruction; the initial delay is fixed at zero. Check identifiability, shifted interval coverage and an independent alignment reference before interpreting the result.

For multishot calls, provide explicit ADC times. The convenience overload has an incorrect time shape, and the per-shot residual operators currently reuse the first shot's time vector. Different shot times with nonzero B₀ are not a supported reliable case. Float64 is the conservative current multishot choice; see its API page for the Float32 default and state-type limitations.

The multishot method estimates one shared scalar, preserving the supplied shot start offsets. Coil compression can reduce repeated encoding work; apply the same fitted transform to both data and CSM as described in coil compression.

Other trajectory-correction methods, including spiral delay calibration, RING and trajectory auto-correction, address related errors with different models and are not interchangeable implementations of this estimator. [9] [10] [11]

Phase preparation checklist ​

Before constructing an encoding operator, check that:

  1. coefficients have been integrated exactly once;
  2. phase in radians has been converted to cycles;
  3. spatial coefficient units match the grid coordinate unit;
  4. coefficient ordering matches the solid-harmonic basis;
  5. zeroth-order phase has not been corrected twice in data and coefficients;
  6. density compensation and encoding use consistent trajectory and timing conventions;
  7. data and coil maps use the same coil ordering and any whitening/compression transform.

Keep timing offsets, tolerances, units and model version with the acquisition's configuration. Use independent field/timing information where available; synthetic checks alone do not validate real-data synchronization or GIRF accuracy.